China Releases 2026 Revised Medical Device Adverse Event Report Form: Go-Live Depends on the System Upgrade

EVERBIZ | Original compilation. Please credit the source when republishing.

China releases 2026 revised medical device adverse event report form

Summary

On 17 November 2026, China’s National Center for ADR Monitoring published the Medical Device Adverse Event Report Form (2026 Revision). The revision was organised by the Center and released after review and approval by the National Medical Products Administration (NMPA), together with a Q&A document to help stakeholders understand the changes.

The notice does not mention any change to existing requirements such as reporting deadlines. What has changed is how to fill in the report, not whether or when to report. The new form will take effect together with the upgraded national monitoring system.

What is the report form, and why revise it?

Once a medical device is on the market, problems such as patient injury or product malfunction may occur during use. These events must be reported to the regulator, and the adverse event report form is the standard format for recording and submitting this information.

According to the official notice, the revision aims to meet the needs of medical device vigilance, build a more scientific and efficient mechanism for collecting risk information, and improve the quality and efficiency of reports.

Put simply: in the past, different reporters described the same problem in very different ways. One would write “broken”, another “ruptured”, another “structural failure”, making it hard to aggregate and analyse the data. The new form is designed to make data accurate, understandable and comparable.

What has changed?

The following is based on the Report Form (2026 Revision) and the accompanying Q&A.

1. Two forms merged into one, with fields shown by reporter type

The previous two forms have been merged into a single form. The online system displays different fields depending on the reporter: registrants and filing holders, distributors, medical institutions, and individual reporters. For example, the firmware version field is only visible to registrants, filing holders and their agents.

2. More detailed information for active and implantable devices

  • Active devices: new fields such as software version and firmware version;
  • Implantable devices: the system automatically calculates the implantation duration from the implantation and explant dates. If the dates are incomplete, the reporter enters the duration directly.

3. IMDRF terminology introduced – the biggest change

The International Medical Device Regulators Forum (IMDRF) maintains a set of adverse event terms that can be seen as a global “standard dictionary” for describing health effects, clinical signs, device problems and the components involved. The new form requires registrants, filing holders and their agents to use these terms in the following places, and the system supports term search:

What is describedRequirement in the new form
Harm to the patientIf “harm” is selected, the clinical signs/symptoms and the health impact must be entered using IMDRF Annex E and F terms
Device problem and component involvedMandatory when the event is classified as a “device problem without harm”; device problems use Annex A terms and components use Annex G terms
InvestigationInvestigation type, findings and conclusions can all be selected from structured terms, including a “to be determined” option; free-text explanations and attachments can also be added

Hospitals and individual reporters can still describe events in free text. If a hospital enters free text or selects a wrong term, the standard term can be added or corrected during the subsequent evaluation.

4. Separate initial and final evaluations

When a registrant or filing holder submits an evaluation, the system asks whether further updates are needed:

  • Updates needed: handled as an initial evaluation, and initial control measures must be entered;
  • Final evaluation, or no further updates needed: final control measures must be entered.

Both stages ask a key question: if changes are made to the design, manufacturing, software, labelling or instructions for use, do these changes require a registration or filing change? In other words, adverse event handling and registration changes are now considered together in the same form.

5. Automatic pre-filling and automatic validation

According to the Q&A, the new system can automatically pull product information from the registration certificate number, filing number or UDI. If mandatory fields are missing, code formats are wrong or dates are inconsistent, the system will block the submission. The data entered can also be reused in periodic risk evaluation reports and in periodic safety update reports under the medical device vigilance pilot.

Active medical device in a testing laboratory

What does this mean for manufacturers?

  • Terminology becomes a core skill. A simple “product damaged” will no longer do; problems must be mapped to standard IMDRF terms. Companies should review internal complaint, after-sales and quality records against these terms in advance.
  • Hospital descriptions must be “translated”. Medical institutions usually report in free text, and registrants need to convert it into standard terms during evaluation. Staff responsible for this need training.
  • Investigations can be staged. The new form allows an “investigation ongoing” status: initial measures first, final conclusions later. Internal adverse event procedures should support this “report first, complete later” approach.
  • Link corrective actions to registration changes. The report must state clearly whether corrective actions involve a registration or filing change, so regulatory and registration teams need to be involved earlier.
  • Master data must be accurate. If registration numbers, UDI, software and firmware versions or implant records are incomplete, the system is likely to block the submission.

Our recommendations

  1. Download the report form and the Q&A from the National Center for ADR Monitoring website and review them item by item against your product types;
  2. Watch for the launch date of the upgraded monitoring system and any follow-up completion guidelines;
  3. Pick one or two past adverse event cases and do a “trial run” with the new form to identify missing information and process gaps early.

EVERBIZ continuously tracks global medical device regulatory developments. If you have questions about China NMPA, EU MDR or Southeast Asian registrations, please contact us.